Electrochemical analysis of gentamicin sulfate

2024
0 views
0 downloads
Advisor: Doç. Dr. Berna Koçak ; Doç. Dr. Yeliz İpek

Abstract (EN)

In this thesis study, for the first time, chromium (III) oxide-glutardialdehyde modified boron doped diamond electrode (Cr2O3-GA2/BDDE) was used as a sensor for the electrochemical analysis of gentamicin sulfate. The Cr2O3-GA2 suspension used in electrode modification was physically coated on the BDD electrode by a simple drop casting method. Characterization of Cr2O3-GA2 used in the modification was made by scanning electron microscopy, energy dispersion spectroscopy (SEM-EDS) and electrochemical method. The behavior of gentamicin sulfate on Cr2O3-GA2/BDDE was done by cyclic voltammetry (CV), and its electrochemical analysis was done by differential pulse voltammetry (DPV). Electrochemical analysis of gentamicin sulfate on Cr2O3-GA2/BDDE was carried out in pH 9 Britton-Robinson (BR) buffer by DPV method. The linear working range of gentamicin sulfate on Cr2O3-GA2/BDDE was calculated as 5.96x10-6- 6.54x10-5 M, and the correlation coefficient was calculated as 0.997. The limit of detection (LOD) was found to be 2.84x10-6 M. Additionally, Genta®, a 3% eye/ear drop solution, was preferred to determine the applicability of gentamicin sulfate in commercial formulation. As a result, it was shown that the method developed on Cr2O3-GA2/BDDE for the determination of gentamicin sulfate can be applied to commercial drugs.

Author

Dr. Osman Ekşi

How to Cite

Osman Ekşi (Master Thesis). Electrochemical analysis of gentamicin sulfate, 2024, Munzur University.