Abstract (EN)
In this study the performance characteristics of all solid-state PVC membrane K+-selective sensors modified with carbon nanotube were investigated. The performances of prepared microsensors were tested by static media measurements. The developed microsensor had a slope of approximately 59.4 mV/ per decade, with a short response time (<15 s) and a long lifetime (4-6 weeks). The developed sensor exhibited a linear behavior in the concentration range of 10-7-10-1 mol.L-1 with a detection limit of 6x10-8 mol.L-1. The determination of potassium in various commercial water samples was performed successfully using the prepared microsensor. The results were compared with the AAS method.
Author
Hande Kaplan
How to Cite
Hande Kaplan (Master Thesis). Development of performance characteristics of potassium selective sensor modified with carbon nanotube, 2019, Giresun University.