Master'sOpen Access

Preparation of a new biosensor for determination of glucose with immobilization on the polypyrrole film of glucose oxidase in biologically fluids

2012
0 views
0 downloads
Advisor: Yrd. Doç. Dr. Servet Çete

Abstract (EN)

In this study, new amperometric enzyme electrode for determination of glucose oxidase was developed. Glucose oxidase enzyme has been immobilized at media prepared with paratoluensulphonate of pyrrole in polypyrrole-paratoluensulphonate film crosslinking via glutaraldehyde method. For this reason; optimum conditions of Pt/ polypyrrole-paratoluensulphonate electrode were determined. Glucose detection is based on the oxidation of hydrogen peroxide at 0,30 V produced by the enzymatic reaction on the enzyme electrode surface. The effects of pH and temperature on the response of the glucose enzyme electrode were investigated. The immobilized of glucose oxidase KM(app.) and Vmax(app.) values were found. Determination of glucose was carried out in biological fluid (in blood) by enzyme electrode. Interference effects were investigated on the amperometric response of the enzyme electrode .Key Words :Glucose, glucose oxidase, enzyme electrode, biological fluid, polypyrrole, paratoluensulphonate

Author

Özgür Bal

How to Cite

Özgür Bal (Master Thesis). Preparation of a new biosensor for determination of glucose with immobilization on the polypyrrole film of glucose oxidase in biologically fluids, 2012, Gazi University.

Keywords

License

Tüm Hakları Saklıdır

This work is shared under the specified license terms.