Master'sOpen Access

Tek anten kullanan kablosuz ve pasif çoklu sensör platformu

2022
0 views
0 downloads
Advisor: Assist. Prof. Dr. Levent Beker

Abstract (EN)

Near field resonant inductive coupling (NRIC) is one of the oldest and predominant wireless communication methods with an implanted medical device (IMD). In this approach, an inductor coupled with a sensor makes an LC resonator. Any change in the sensor's output is reflected as a change in its capacitance, which in turn leads to a shift in this LC resonator's resonance frequency. Shift in this resonance frequency can be tracked wirelessly by measuring the output impedance/reflection coefficient of an external reader antenna in the near field of the implanted device. In most current devices, data from multiple sensors is required simultaneously. In this work, we propose a simple configuration to simultaneously sense various parameters from one antenna. By tapping the inductor at different locations, we get two sub-inductors. We connect two capacitive sensors with the sub-inductors to get two LC loops. The scope of this thesis is the design of such an antenna through simulations, its fabrication, and characterization with two sensors simultaneously to demonstrate its feasibility in potential IMD applications.

Author

Dr. Muhammad Junaıd Akhtar

How to Cite

Muhammad Junaıd Akhtar (Master Thesis). Tek anten kullanan kablosuz ve pasif çoklu sensör platformu, 2022, Koç University.

License

Tüm Hakları Saklıdır

This work is shared under the specified license terms.

More theses from Koç University