Master'sOpen Access

Reduction behaviour of moxifloxacine hydrochloride and its determination in human urine and tablet dosage form by voltammetry

Is this your thesis?

This record came from a bulk archive import. If it’s yours, link it to your profile.

2018
0 views
0 downloads

Abstract (EN)

A voltammetric method was used for the quantitative determination of moxifloxacin hyrochloride (MOXHCl) in pharmaceutical preparations and spiked human urine. The electrochemical experiments were carried out in different buffer solutions in the pH range from 0.51 to 12.00 on glassy carbon electrode (GCE) by cyclic voltammetry (CV) and differential pulse voltammetry (DPV) techniques. The effect of pH on the peak current and peak potential was investigated. Phosphate buffer (pH 6.50) was selected for analytical purposes. The peak was established to be diffusion controlled nature of electrode. A linear calibration curve for DPV analysis was constructed in the MOXHCl range from 3x10-4 M to 1.1x10-3M. Limit of detection (LOD) and quantification (LOQ) were obtained as 5.51x10-5 M and 2.30x10-4 respectively. Applied method was valided.

Author

Burcu Tepeli

How to Cite

Burcu Tepeli (Master Thesis). Reduction behaviour of moxifloxacine hydrochloride and its determination in human urine and tablet dosage form by voltammetry, 2018, Çanakkale Onsekiz Mart University.

Keywords

License

Tüm Hakları Saklıdır

This work is shared under the specified license terms.

More theses from Çanakkale Onsekiz Mart University