Investigation of dosimetric and thermoluminescence properties of CaF2:Mn (TLD-400)
Bu tez size mi ait?
Bu kayıt toplu arşivden geldi. Sizinse profilinize bağlayın.
Özet (EN)
The additive dose (AD), Tm(Ea)?Tstop, repeated initial rise (RIR), peak shape (PS), and computerized glow curve deconvolution (CGCD) methods were used to analyze the thermoluminescent (TL) dosimetric glow peak of CaF2:Mn after ß- irradiation between the dose level D=0.015 Gy and D=110 Gy. The activation energies found with the IR and PS methods yield very close values after the cleaning of low temperature satellite peaks from the dosimetric glow peak by increasing stopping temperature in the Tm?Tstop experiment. The influence of heating rates on the response of dosimetric glow peak of CaF2:Mn was also studied by the peak height and peak area methods using CGCD technique. It was observed that the peak intensities and total area of glow peak decrease continuously with increasing heating rate.Another dosimetric characteristic which is the dose response of this peak was also examined by peak height and area methods. Keywords: Thermoluminescence; CaF2:Mn; TLD-400; Kinetic parameters; Dose response; Heating rate.
Yazar
Ayşe Sibel Sökücü
Bu Yayına Nasıl Atıf Yapılır
Ayşe Sibel Sökücü (Master Thesis). Investigation of dosimetric and thermoluminescence properties of CaF2:Mn (TLD-400), 2007, Gaziantep University.
Anahtar Kelimeler
Lisans
Tüm Hakları Saklıdır
Bu eser belirtilen lisans koşulları altında paylaşılmaktadır.
Gaziantep University tezlerinden daha fazlası
- Police profession and power to use force(2023)
- Chronic effect of jumping rope exercises on inspiratory and expiratory respiratory muscles(2024)
- Pilton (pastinaca armena) katkılı beyaz peynirin duyusal ve kimyasal özelliklerinin incelenmesi(2019)
- Analysis of tooth stiffness of spur gears by finite element method(2017)
- Constructions of popular culture within viral advertising: Reception analysis of Eti Benim'O virals(2021)
- Optimum usage of mixed damping systems (rubber concerete or x diagonal dampers) on multystory building(2021)