The evaluation of validity a portable clinical analyzer for the determination of blood ionized calcium concentration in cattle
2018
0 views
0 downloads
Advisor: Prof. Dr. Tolga Karapınar
Abstract (EN)
Information on reliability of the i-STAT, a point-of-care (POC) device, to determine blood ionized calcium (iCa2+) concentration is limited. Rapid determination of blood iCa2+ concentration can help evaluation of calcium status and delivery of effective therapy in field conditions. The goal of this study was to evaluate reliability of the i-STAT (Abbott Point of Care, Abbott Laboratories, Chicago, IL) in measuring blood iCa2+ concentration in cattle. In this prospective study 121 cattle with various diseases were used. Venous blood samples were run the i-STAT and the Radiometer ABL 800 (Radiometer, Copenhagen, Denmark) analyzer for determination of blood iCa2+ concentration. Data were subjected to paired Student t-test, concordance analysis, Passing-Bablok regression and Bland-Altman plots for comparison, accuracy, agreement and reliability. The Radiometer ABL 800 was reference method however the i-STAT was the test method. The i-STAT performance was also evaluated at two different cut-off values for blood iCa2+ concentration (< 1 mmol/L; <1.18 mmol/L) using the receiver operating characteristics (ROC) curve. The mean blood iCa2+ concentration measured by the i-STAT was similar to the mean blood iCa2+ concentration measured by the Radiometer ABL 800 (1.09 versus 1.09 mmol/L; P = .82). The i-STAT results were highly correlated with the Radiometer ABL 800 results (r= 0.98, P < .0001). Regression equation fitting blood iCa2+ concentration measured by the i-STAT (Y) and blood iCa2+ concentration measured by the Radiometer ABL 800 (X), was Y= -0.108276+1.103448× X, with a residual deviation (RSD) of 0.02. The deviation from the identity line was not significant (P = .64). The Bland-Altman difference plot revealed that the mean bias was zero. The i-STAT had Se of 100 % (95% CI, 87,7 - 100,0) and Se of % 94.8 (95 % CI, 87.2-98.6) at the cut-off blood iCa2+ concentration < 1 and < 1.18 mmol/L, respectively. The i-STAT yielded precise and accurate blood iCa2+ results in cattle. The i-STAT especially predicted hypocalcemia with high sensitivity. Key Words: calcium, dairy cow, point-of-care (POC), sensitivity.
Author
Öznur Yılmaz
How to Cite
Öznur Yılmaz (Master Thesis). The evaluation of validity a portable clinical analyzer for the determination of blood ionized calcium concentration in cattle, 2018, Fırat University.
Keywords
License
Tüm Hakları Saklıdır
This work is shared under the specified license terms.
More theses from Fırat University
- Using social media as an integrated marketing communication tool(2018)
- Foundation of Dutch East İndia Company and her rising in İndonesia in the 17th century(2013)
- Examination of stress state between Doğanyol (Malatya) and Çelikhan (Adıyaman) on the east Anatolian fault zone(2020)
- Color usage at Turkish Divan of Fuzûlî(2013)
- Yavuzeli (Gaziantep) surrounding volcanic outcropping of rocks petrographic and geochemical features(2014)
- Hizbu?t-Tahrir and the religions and political thoughts of Ercumend Özkan(2008)