The comparison of photon dosimetries with different ion chambers in linear accelerator used in radio theraphy
2009
0 görüntülenme
0 i̇ndirme
Danışman: Doç. Dr. Eyüp Tel
Özet (EN)
In this study Saturne-43 Linear Accelerator treatment machine?s 6, 15 and 25 MV photon beams, and different ion chambers at Dr. Abdurrahman Yurtarslan Ankara Oncology Education and Research Hospital were used. Measurements were made using liquid and solid phantoms. Dose efficiency measurement was made in 5 cm depth for 6 MV energy photons and in 10 cm depth for 15 MV and 25 MV energy photons. In middle size water phantom measurements were made in the same depths. Benefiting from these measurements TPR20 10, M+, M-, Q1, Q2 and dose efficiency values were obtained for different ion chambers. With the data obtained from these measurements, calculations were made according to IAEA Technical Report Series No.277. Dosimetries calculated for different energy and ion chambers were compared.
Yazar
Sümeyra Mandal
Bu Yayına Nasıl Atıf Yapılır
Sümeyra Mandal (Master Thesis). The comparison of photon dosimetries with different ion chambers in linear accelerator used in radio theraphy, 2009, Gazi University, Fizik Bölümü.
Anahtar Kelimeler
Lisans
Tüm Hakları Saklıdır
Bu eser belirtilen lisans koşulları altında paylaşılmaktadır.
Gazi University tezlerinden daha fazlası
- Occupational accident analysis and modelling in oil and gas drilling sector Turkey(2021)
- XVI. yüzyıl Anadolu'sunda Oğuzların Karkın Boyu(2004)
- Sharing of real life geometry samples via a social learning environment: A case study(2021)
- Evaluatıon of calcium hydroxide removal efficiency of two different irrigation activation techniques from artificial internal resorption cavities prepared at different root levels(2021)
- Experimental development of the interfacial bond-slip model between textile reinforced mortar strips and masonry walls(2025)
- The use of verbal memory in the context of sustainability and power at the museums of Turk(2010)
