Master'sOpen Access

Development of potentiometric isoprenaline- selective electrode and its applications

Is this your thesis?

This record came from a bulk archive import. If it’s yours, link it to your profile.

2019
0 views
0 downloads

Abstract (EN)

In this study, a potentiometric ion-selective electrode (ISE) was used to determine isoprenaline, a pharmaceutical active substance. IP-FM, IP-TSA, IP-TPB, IP-REY, IP-FTA ion pairs were synthesized. The synthesized ion pairs were used as the ionophore in the structure of the electrode membrane. In order to ensure membrane optimization, PVC membrane ion selective electrodes were produced in various compositions using the synthesized ion pair and the potentiometric performance characteristics of these electrodes were investigated. It was determined that the best potentiometric performance characteristics were 3.0% isoprenaline-tetrafenylborate ion pair, 65.0% dibutylphthalate, 32.0% polyvinylchloride and % 1 Tetrakis (4-chlorophenyl) boron potassium compound. The linear operating range of this electrode is 5.0x 10-6 M - 1.0x 10-1 M and the slope at the 10-fold concentration change is 45.3 mV; determination limit, 5.0x10-6 M, pH working range, 2.6-3.6 and 5.7-7.9; response time, < 5s. The electrode exhibited a reproducible potentiometric response. Isoprenaline content was determined in drug by using the isoprenaline selective electrode.

Author

Oktay Ozarslan

How to Cite

Oktay Ozarslan (Master Thesis). Development of potentiometric isoprenaline- selective electrode and its applications, 2019, Van Yüzüncü Yıl Üniversity.