Determination of lead and cadmium in textile leach solutions using chemical vapour generation atomic absorbtion spectrometry
2004
0 görüntülenme
0 i̇ndirme
Danışman: Y.doç.dr. Nusret Ertaş
Özet (EN)
A simple and rapid method using flow injection cold vapour AAS and hydride generation AAS were described for the determination of cadmium and lead in textiles leach solutions, respectively. The experimental conditions were optimized for the best sensitivities. The detection limits (3 a) in leachate solutions are 0.05 ng/mL for cadmium and 1.0 ng/mL for lead with a precision between 1- 3% RSD. The accuracy of methods were tested using ETAAS as a comparison method and a standard reference material. In both cases, the results were in good agreement at 95% confidence level. Textile products obtained from local markets were leached in artificial sweat solution. The concentrations of cadmium and lead in textile products were found to be below the limit values assigned by Oeko-tex standards.
Yazar
Dr. Orkun Alp
Bu Yayına Nasıl Atıf Yapılır
Orkun Alp (Master Thesis). Determination of lead and cadmium in textile leach solutions using chemical vapour generation atomic absorbtion spectrometry, 2004, Gazi University.
Anahtar Kelimeler
Lisans
Tüm Hakları Saklıdır
Bu eser belirtilen lisans koşulları altında paylaşılmaktadır.
Gazi University tezlerinden daha fazlası
- Occupational accident analysis and modelling in oil and gas drilling sector Turkey(2021)
- Experimental development of the interfacial bond-slip model between textile reinforced mortar strips and masonry walls(2025)
- Deveplopment of semiconductor humidity sensors(2021)
- The effect of computer-assisted and direct strategy teaching on reading comprehension(2021)
- Investigation of engineer applicant students' conceptual and procedural knowledge of integral problem solving process from the point of bloom taxonomy(2021)
- TCDD (The State Railways of the Republic of Turkey)`s state ve financial structure(2006)