A study on influence the phantom dimensions of the dose values around of the I-125 brachytherapy sources by using Monte Carlo technique
Is this your thesis?
This record came from a bulk archive import. If it’s yours, link it to your profile.
Abstract (EN)
In this study for 125I brachytherapy seed sources, the influence of phantom dimensions ondose values were calculated by using a number of cubic water phantoms with variable sidelengths. In each calculation, the source was positioned at the center of the phantom. Two differentseed models were used. Namely, Amersham 6711 OncoSeed and Imagyn IsoSTAR. In addition,the source was translated from 1 to 4 cm along both long axis and transverse axis of the source inorder to calculate the variation of dose at selected points in the phantom.Dose values were calculated by using EGSnrc Monte Carlo code. The geometries of 125Iseeds and surrounding medium were modeled by using Yegin's multi-geometry package.125Keywords : Brachytherapy, Multi-geometry teknique, EGSnrc, I, Monte Carlo.
Author
Sebile Duman
How to Cite
Sebile Duman (Master Thesis). A study on influence the phantom dimensions of the dose values around of the I-125 brachytherapy sources by using Monte Carlo technique, 2007, Manisa Celal Bayar University.
Keywords
License
Tüm Hakları Saklıdır
This work is shared under the specified license terms.
More theses from Manisa Celal Bayar University
- Analysing the social integration process of yezidi refugee youngs in the context of multiculturalist social work(2016)
- Anomie in university students: MCBU case(2023)
- Türk dış politikasında kriz yönetimine örnek bir vaka: Struma olayı(2025)
- TÜRK FİKİR HAYATINDA MİLLİYETÇİ MUHAFAZAKAR KADIN FİGÜRÜ: AYŞE DERGİSİ, EMİNE IŞINSU VE SAMİHA AYVERDİ(2025)
- Identity perceptions of Izmir Jews(2019)
- Ottoman Singapore relations in the reign of second Abdulhamid (1876-1908)(2019)