Evaluation of scintigraphic images of simulated tumors with different volumes in a liver phantom
Is this your thesis?
This record came from a bulk archive import. If it’s yours, link it to your profile.
Abstract (EN)
Objective: Yttrium-90 (⁹⁰Y) radioembolization and the use of Technetium-99m macroaggregated albumin (⁹⁹ᵐTc-MAA) for pre-dosimetry are routinely effective methods in the treatment of unresectable liver tumors. This study aims to evaluate the impact of tumor volume variation and tumor proximity on scintigraphic image quality and dosimetry, and to objectively analyze these effects using contrast and contrast-to-noise ratio (CNR) parameters. Materials and Methods: In this experimental study, a specially designed liver phantom containing four simulated tumors with diameters of 1 cm, 2 cm, 3 cm, and 5 cm was used. The tumors and the surrounding liver parenchyma-mimicking areas were filled with ⁹⁹ᵐTc at a ratio of 4.88:1. The phantom was imaged in a water tank using standard clinical protocols for both planar and SPECT scintigraphy. Regions of interest (ROIs) were drawn over the lesions on the images to calculate contrast, CNR, and tumor doses. Results: In SPECT images, as tumor diameter increased, tumor-to-normal liver ratio (T/N) per milliliter, contrast, CNR, and tumor doses also increased. Tumor doses for 1 cm and 5 cm diameter lesions were found to be 6.9 Gy/ml and 13.43 Gy/ml, respectively. Similarly, T/N ratios, contrast, and CNR values calculated from planar imaging also increased with tumor size; however, the increase in tumor doses was found to be relatively lower. Discussion and Conclusion: This study analyzed the mutual effect of ROI counts in the presence of small and multiple tumors in ⁹⁹ᵐTc-MAA pre-dosimetry for liver tumors. It was observed that smaller tumors (1–3 cm) located near larger tumors were affected by spillover from the larger lesions, resulting in artificially elevated ROI counts and consequently overestimated dose calculations. In conclusion, the presence of larger lesions may influence dosimetric evaluations of smaller (1–3 cm) tumors, potentially leading to inaccurate dose assessments. Keywords: ⁹⁹ᵐTc-MAA scintigraphy, ⁹⁰Y radioembolization, pre-dosimetry, image quality, tumor doses
Author
Murat Kozanlılar
Institution
How to Cite
Murat Kozanlılar (Master Thesis). Evaluation of scintigraphic images of simulated tumors with different volumes in a liver phantom, 2025, İstanbul University-Cerrahpaşa.
Keywords
License
Tüm Hakları Saklıdır
This work is shared under the specified license terms.
More theses from İstanbul University-Cerrahpaşa
- Reinterpreting the street as a field of work from the perspective of social work in türki̇ye: A field study in the case of istanbul(2024)
- Developing 3D printer for thermoelectrical material and module fabrication(2024)
- The effect of mentoring programme on reality shock, clinical adaptation and professional competencies of new graduated nurses(2024)
- Evaluation of the effectiveness of safe nutrition program applied to acute stroke patients with dysphagia(2024)
- Joint inversion of H/V and SPAC methods: A case study of Eskişehir Basin(2024)
- Nurses' views on motivation and development of the nurse motivation scale(2024)